Wetted Materials
The wetted parts of "Standard configuration" Encynova System Modules are made from chemically-resistant materials, creating a versatile, durable fluid control system. For applications that involve more aggressive fluids, an even more robust "Inert" wetted parts option is available. Other options may be available- please contact us about your 'problem liquid.'
  Standard Inert
Housing Delrin (2-series)
Ertalyte or Aluminum (16-series)
PEEK (2-series)
Ertalyte or Aluminum (16-series)
Manifold PEEK, Tefzel, or
316L Stainless Steel
PEEK, Tefzel, or
316L Stainless Steel
Manifold O-rings Viton Perfluoroelastomer
Port Plates Ceramic Al2O3 Ceramic Al2O3
Pistons Teflon or UHMW Teflon or UHMW
Cylinders Borosilicate glass (Pyrex) Borosilicate glass (Pyrex)
Connecting Rods Delrin PEEK
click here for pdf
click here for pdf

Encynova OEM System Modules-

productstechnologytechnical informationcontact us

Encynova System Modules are complete fluid delivery systems for metering and dispensing. These compact modules precisely move, control, and measure fluid.

Nova System Communications
To harness precision, you must have control.

That's why our Nova Systems have a flexible communications platform. RS-485 communication capability allows you to easily control the Nova Systems with our predefined commands. Use a PC, PLC, or any other source that can generate 8-bit ASCII commands to precisely control as many Nova Systems as you need.

LabVIEW™, C++ and Visual Basic compatibility provide the versatility required for simple and effective integration into your process. Connect the Nova System, assign it to an address on your network, and you are ready to go. Up to 250 Modules can be controlled on one RS-485 network using ASCII serial strings. Digital I/O and analog capability are standard in Nova Systems and in Basic Systems that use the integrated motor/driver.

Dispensing specs are based on tests with 1.0 mL of RO water at room temperature and 1 atm of pressure.
* Please consult Encynova for information specific to your application.
* * Specifications are dependent on control system used.
* * * For flow rates of less than 1 mL/min the calculated instantaneous pulsation levels increase.
Observed pulsation levels depend greatly on fluidic setup and measurement technique.

Cv: The coefficient of variation, expressed in percent, is used to describe the standard deviation (s) relative to the mean (X). Cv = [ (100) (s)/ X ]

www.encynova.com ---------1.303.465.4800 --------toll-free 1-866-ENCYNOVA

 

 

2-Series

Model 2-1
Low-cost metering up to 50 ml/min.

Single-cylinder design; pulsation depends on control method.

Model 2-4
Dispensing and low-pulsation metering up to 200 ml/min.

Nova 2-4 System includes NovaFlow™ for near-pulseless metering, and NovaSpense™ for dispensing accuracy up to 0.1% Cv.

All dimensions shown in inches

16-Series

16-series systems are recommended for applications with higher-viscosity fluids.

Model 16-2
Dispensing and metering up to 640 ml/min.

Model 16-4
Dispensing and low-pulsation metering up to 1280 ml/min.

Nova 16-4 System includes NovaFlow™ for near-pulseless metering, and NovaSpense™ for dispensing accuracy up to 0.1% Cv.

Move it. Control it. Measure it.

Encynova systems are different from basic pumps because they do more than typical pumps. Measurement and flow control capabilities are incorporated into the pump's design. No external valves or flow meters are needed.

Specifications

Nova System Modules  Model 2-4  Model 16-4
Metering    
----  Metering Range 200 nl/min - 200 ml/min 1600 nl/min - 1280 ml/min
----  Metering Accuracy < 0.5 %Cv < 0.5 %Cv
NovaFlow™ Metering    
----Range
200 nl/min - 20 ml/min 1600 nl/min - 200 ml/min
----Accuracy
< 1 %Cv < 1 %Cv
----Pulsation (%Cv of total flow at mid range rate)
3 %Cv 3 %Cv
NovaSpense™ Dispensing    
  ---- NovaSpense™ Dispensing Accuracy 0.1 %Cv 0.1 %Cv
  ---- NovaSpense™ Dispensing Precision 0.05 %Cv 0.05 %Cv
----Motor: Servo with features for increased dispensing accuracy-and near-pulseless metering.
Basic System Modules Model 2-1 Model 2-4 Model 16-2 Model 16-4
Metering**        
---- Range w/ stepper motor: integrated motor/driver 0.2 - 50 ml/min .08 - 200 ml/min    
---- Range w/ stepper motor: Two-phase 1.8 degree 0.2 - 50 ml/min 6 - 200 ml/min 25 - 640 ml/min 50 - 1280 ml/min
---- Accuracy < 2 %Cv < 2 %Cv < 2 %Cv < 2 %Cv
Dispensing        
---- Accuracy** 1-20 %Cv 1-20 %Cv 1-20 %Cv 1-20 %Cv
---- Repeatability < 1 %Cv < 1 %Cv < 1 %Cv < 1 %Cv
General specs for all systems Model 2-1 Model 2-4 Model 16-2 Model 16-4
----Number of cylinders 1 4 2 4
----Displacement volume per cylinder (ml/rev) 0.2 0.2 1.6 1.6
----------------------------------------------total (ml/rev) 0.2 0.8 3.2 6.4
----Pressure max. operating 125psi (8bar) 125psi (8bar) 125psi (8bar) 125psi (8bar)
----Manifold Fittings 1/16" NPT 1/16" NPT 1/8" NPT 1/8" NPT
----(custom fittings are available) ¼"-28 UNF ¼"-28 UNF ¼" NPT ¼" NPT
----Approx. Dimensions l x w x h (inches) 5.6 x 2.2 x 2.8 7.6 x 2.8 x 2.8 9.6 x 5.0 x 5.0 9.6 x 5.0 x 5.0
----Weight (varys depending on motor type) 1.6lb (0.7kg) 3.8lb (1.7kg) 8.6lb (3.9kg) 11.1lb (5.0kg)

Four Model 2-1 System Modules
panel-mounted in a standard 19-inch rack.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

top

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

top

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

top

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

top

 

 

 

 

 

 

 

 

 

 

 

 

 

top